| Model Name: |
Pointe 180 II |
| Spectrophotometer type: |
Filter |
| Optical Configuration: |
Single beam with continuously rotating filter wheel. Monochromatic or bichromatic reading 8 filter positions. |
| Usable Spectral Range: |
330 to 770 nm |
| System Procedures: |
Open and Stored Menu. |
| Calculating Modes: |
Absorbance, Single Standard (differential samples) Factor Mode (differential samples) Multi Standard Mode (up to 7 standards) Multi Standard % Mode (up to 7 standards) Kinetic Mode (consecutively, or simultaneously (Batch)) By factor or standard. Fixed Time Kinetic By factor or standard. Index Mode. |
| Channels: |
120 open |
| Source of Radiation: |
Tungsten Halogen, 10 watt, with automatic lamp saver. |
| Selection if Wavelength: |
By Filter. |
| Filter Type: |
4-cavity interference, long-life ion beam-assisted deposition. |
| Wavelength accuracy: |
± 3 nm |
| Filter location: |
After sample (heat absorbing filter before sample). |
| Filter selection: |
Automatic by software or via keyboard. |
| Wavelengths: |
340, 405, 505, 545, 580, 630, nm supplied standard, others optional. |
| Half bandwidth: |
<10 nm |
| 1/100 bandwidth: |
14 nm at 340 nm. |
| False radiant energy ratio: |
<0.001 at 340 and 405 nm. |
| Cuvette: |
1 cm square, 12 mm round, flow through. |
| Type supplied: |
Flow through. |
| Material: |
316 stainless, borosilicate windows. |
| Geometry: |
Cylindrical, 2.3 mm dia x 5 mm +/- 0.05 mm. |
| Illuminated Volume: |
21 ul |
| Minimum read volume: |
250 ul |
| Aspiration / purge: |
Vacuum pump at 18 cm of Hg. |
| Valve: |
Silicone pinch type |
| Cuvette Holder: |
Thermostatically controlled compartment at 37°C. |
| Detector: |
Gallium-Aresnide-Phoshide photoiode. |
| Display type: |
240x128 graphic LCD w/backlight |
|
| Scale of Display: |
|
| Absorbance: |
-0.5 to 3.5 (flow through mode) -0.5 to 2.5 (tube or 1 cm cuvette) |
| Concentration: |
Maximum 999,999 |
| Kinetic Results: |
Abs/min with resolution of 0.0002 A/min |
| Zero Compensation: |
Automatic |
| Range: |
-0.5 to 2.5 absorbance |
|
| Signal Outputs: |
| Parallel: |
Centronics/IBM-PC compatible |
| Serial: |
RS-232 at 9600 baud, 8 data, 1 stop, no parity Bi-directional. |
|
| Data input: |
1) 20 key keypad, 2) PS2 101 keyboard (connector on back) |
| Spectrophotometric Inaccuracy Flow-through: |
<0.5 % at 1 absorbance, 340630 nm NADH Solution. <0.5 % at 2 absorbance, 340/630 nm NADH Solution. <0.5 % at 3 absorbance, 340/630 nm NADH Solution. <0.5 % at 1 absorbance, 405/630 nm PNP Solution. <0.5 % at 2 absorbance, 405/630 nm PNP Solution. <0.5 % at 2 absorbance, 405/630 nm PNP Solution. |
| Stability: |
Better than 0.003 A/hr monochromatic after warmup. |
| Warmup time: |
90 seconds photometric, 15 minutes for temperature compartment. |
| Electronics: |
Z180 Microprocessor 18 MHz 128k EPROM. |
| Power: |
Auto-Switching Power supply. Voltage source: 90-264 VAC Frequency: 50/60 Hz Power consumption: 60 watts Installation Category: CAT II Fuses: 2.5A/250V fast 5-20 mm glass fuse, 2.0A/250V Fast 5-20 mm Ceramic Fuse, (2) 6/10 250V slow blow 3AG Fuses
|
| Dimensions and Weight: |
40cm (L) x 37 cm (W) x 14 cm (H) lid closed (30 cm lid opened), 6.4 kg |
| Space Requirements: |
10 cm clearance on all sides. |
| Environmental Conditions for Safe operation: |
Indoor Use Altitude up to 2000m. Temperature 5°C to 40°C (Although it may be safe to operate in these conditions, it may not be suitable for the performance of the user’s tests. Humidity 85% for temperatures up to 30°C decreasing linearity to 50% humidity at 40°C. Mains supply voltage fluctuations not to exceed +/- 10% of nominal voltage.
|
| Recommended Operating Temperature: |
15-35°C
|
| Recommended Operating Humidity: |
Between 10 and 85%, non-condensing
|