Photometric:
Linear measurement range: 0.00 to 3.0 Absorbance Units (A)
Photometric accuracy: ±(1% of the reading + 0.01 A) from 0 to 1.5A
±(2% of the reading + 0.01 A) at 1.5-3.0A
Stability: drift of no more than 0.005A in 8 hours
Warm up Time: 45 seconds, built-in
Light Source: Tungsten Lamp
Standard Wavelengths: Standard: 405, 450, 492, and 630nm
Type of filter: 3-cavity hard coat interference filter, 10nm half bandpass typical
Vessel: single, double or break apart strips, up to 12 wells long 3 strip load capacity
Electronic and Software:
Speed: reads, calculates and prints results for up to 12 wells in about 30 seconds.
Display: LED, 16 segment alphanumeric
Internal Printer: thermal, dot matrix, 20 characters per line
Keyboard: 16 key, membrane switch, 4X4
Calculation Modes: single point calibration, multipoint calibration with regression or point to point curve fit, cutoff, % Absorbance Multi-point, Uptake
Serial Port: output only, 2400 Baud, 1 start bit, 8 data, 1 stop, no parity, no handshake
Power Source: Switch selectable power supply (115V or 230V indicated)
Voltage Source 110-120/220-240 VAC from 50 to 60 Hz, CAT II
Power consumption is less than 50 Watts
Fuse: two 1/2A, T rating, 250V
All power cords must be approved for the country of use.
Enclosure: ABS Plastic cover with metal base
Dimensions: Approx. 9” X 12” X 3”, (23 X 30 X 8 cm)
Weight: approx. 13 lbs. (5.9 Kg.)
Other:
Certifications & Compliances: Listed to Standard UL 1262, Laboratory Equipment<br>CE Certificate of Conformity. Conforms to the following standards: EN 50082-1, EN 55022, En 61010-1 following the provisions of the 73/23/EEC and 89/336/EEC
Environmental Conditions for safe Operation: Indoor use Altitude up to 2000m Temperature 5°C to 40°C (Although it may be safe to operate in these conditions, it may not be suitable for the performance of your tests; check with your supplier.) Humidity 80% for temperature up to 31°C, decreasing linearity to 50% humidity at 40°C Mains supply voltage fluctuations not to exceed ± 10% of nominal voltage.
Optional Accessories: DRI-DYE Check Strips for instrument QC, replacement lamps, fuses, thermal paper, serial cable

  