| Photometric: |
| Linear measurement range: |
0.00 to 3.0 Absorbance Units (A) |
| Photometric accuracy: |
±(1% of the reading + 0.01 A) from 0 to 1.5A ±(2% of the reading + 0.01 A) at 1.5-3.0A |
| Stability: |
drift of no more than 0.005A in 8 hours |
| Warm up Time: |
45 seconds, built-in |
| Light Source: |
Tungsten Lamp |
| Standard Wavelengths: |
Standard: 405, 450, 492, and 630nm |
| Type of filter: |
3-cavity hard coat interference filter, 10nm half bandpass typical |
| Vessel: |
single, double or break apart strips, up to 12 wells long 3 strip load capacity |
| Electronic and Software: |
| Speed: |
reads, calculates and prints results for up to 12 wells in about 30 seconds. |
| Display: |
LED, 16 segment alphanumeric |
| Internal Printer: |
thermal, dot matrix, 20 characters per line |
| Keyboard: |
16 key, membrane switch, 4X4 |
| Calculation Modes: |
single point calibration, multipoint calibration with regression or point to point curve fit, cutoff, % Absorbance Multi-point, Uptake |
| Serial Port: |
output only, 2400 Baud, 1 start bit, 8 data, 1 stop, no parity, no handshake |
| Power Source: |
Switch selectable power supply (115V or 230V indicated) Voltage Source 110-120/220-240 VAC from 50 to 60 Hz, CAT II Power consumption is less than 50 Watts Fuse: two 1/2A, T rating, 250V All power cords must be approved for the country of use. |
| Enclosure: |
ABS Plastic cover with metal base |
| Dimensions: |
Approx. 9” X 12” X 3”, (23 X 30 X 8 cm) |
| Weight: |
approx. 13 lbs. (5.9 Kg.) |
| Other: |
| Certifications & Compliances: |
Listed to Standard UL 1262, Laboratory Equipment<br>CE Certificate of Conformity. Conforms to the following standards: EN 50082-1, EN 55022, En 61010-1 following the provisions of the 73/23/EEC and 89/336/EEC |
| Environmental Conditions for safe Operation: |
Indoor use Altitude up to 2000m Temperature 5°C to 40°C (Although it may be safe to operate in these conditions, it may not be suitable for the performance of your tests; check with your supplier.) Humidity 80% for temperature up to 31°C, decreasing linearity to 50% humidity at 40°C Mains supply voltage fluctuations not to exceed ± 10% of nominal voltage. |
| Optional Accessories: |
DRI-DYE Check Strips for instrument QC, replacement lamps, fuses, thermal paper, serial cable |