Glucose Oxidase

Catalog #: 15-G7521-152


METHOD

Oxidase/Trinder, Endpoint

TECHNICAL INFORMATION

Format: Liquid
Wavelength: 500 nm
Linearity: 0 - 500 mg/dL
Expected Values: 70-105 mg/dL
Storage Temperature: 2-8°C
Reconstituted Stability: Ready to use liquid
Configuration: 4 x 38 mL
Test/Kit: 1,013
Disclaimer: Actual test/kit numbers may be lower due to instrument limitation.
  • Description
  • Ordering Information
  • Related Products
  • Documentation
Intended Use
For the quantitative determination of Glucose in serum on the Tokyo Boeki Medisys Inc. Biolis 24i analyzer. For in vitro diagnostic use only.
Principle
Glucose is oxidized by glucose oxidase to gluconate and hydrogen peroxide. Phenol + 4-AAP + hydrogen peroxide, in the presence of peroxidase, produces a quinoneimine dye that is measured at 500nm. The absorbance at 500nm is proportional to the concentration of glucose in the sample.