US Federal Contractor Registration
CE Marked

Microalbumin

Catalog #: 15-M7562-64


METHOD

Immunoturbidimetric

TECHNICAL INFORMATION

Format: Liquid
Wavelength: 340/700 nm
Linearity: 0.5-30.0 mg/dL
Expected Values: 30-300 mg / 24 hours
Storage Temperature: 2-8°C
Reconstituted Stability: Ready to use liquid
Configuration: 2 x 24 mL / 2 x 8 mL
Test/Kit: 320
Disclaimer: Actual test/kit numbers may be lower due to instrument limitation.
  • Description
  • Ordering Information
  • Related Products
  • Documentation
Intended Use
For the quantitative determination of low levels of albumin in urine by immunoturbidimetric assay on the Tokyo Boeki Medisys Inc. Biolis 24i analyzer. For in vitro diagnostic use.
Principle
When a sample is mixed with anti-human albumin goat antiserum, agglutination is caused by the antigen-antibody reaction. The turbidity is measured at 340nm and 700nm and albumin in the sample is quantitatively determined.